12.4
ELECTRICAL CONNECTIONS
Warning
Notice
The CSS-VC.. is to be supplied in voltages 230 V / 50 Hz or 115 V / 50-60 Hz (for the circuit diagram
see appendix). Protection is provided as standard with two 6.3 A fuses. These are located in the cold-
device socket.
The electrical connection is via the 2 m [≈ 6.56 ft] long power cord with cold-device connector depending
on the mounting of the connection plate either at the back or in the lid of the housing (see Section 12.2).
12.4.1 GROUP CONNECTION ALARM (PART NO. 01G6170)
The electrical connection of the general alarm (condenser, fluid and flow) is via a 6-pin connector,
depending on the mounting of the connection plate, either at the back or in the lid of the housing.
The corresponding 6-pin socket is included.
The connector is wired as follows:
Internal
Lower channel
Channel 1
Upper Channel
Channel 2
Figure 6
Plug assignment for the version with group alarm (Part No. 01G6170)
22
Incorrect voltage can destroy the device. On connection, ensure the
correct mains voltage in accordance with the rating plate!
For the erection of power installations with nominal voltages up to
1000 V the requirements of VDE 0100 and relevant standards and
regulations must be complied with!
Status Alarm
Socket/Female X2
External
6
5
4
3
2
1
CSS-VC-1.6 | 1.01.03
Contact load
Contact rating
24V AC/DC 0,5A
Plug/male connector type Phoenix MSTB
2,5/6-STF 5,08
www.mc-techgroup.com