Cooler Controller - M&C CSS Series Instruction Manual

Gas conditioning unit, 19“, wall mount or portable in carrying case with software version 1.6
Hide thumbs Also See for CSS Series:
Table of Contents

Advertisement

The following steps are to be carried out before initial commissioning:
Plug the cold-device connector of the supplied mains connection cable into the cold-device socket
for gas conditioning;
Connect power cord to the network;
Switch on power supply.
After the end of the lead time, the gas conditioning unit is ready for operation (green LED on the
controller lights up).
Notice
13.1

COOLER CONTROLLER

Display
> 8 °C
After
[> 46.4 °F]
commissioning
(room
temperature)
After about. 10
min.
[≤ 41.9 °F]
Normal operation 5 °C [41 °F]
Notice
26
In the case of long-term measurements with high dust content in the
sample gas, a suitable gas sampling probe to protect the sample line
from clogging must be provided.
LED red
ON (status
alarm)
5.5 °C
OFF
OFF
If a cooler temperature alarm is present, after default setting, the gas
sample pumps are switched off (if available) to prevent the transport of
wet sample gas into the analyzer.
CSS-VC-1.6 | 1.01.03
LED yellow
LED green
Lights up in the
case of an existing
OFF
flow alarm
OFF
ON (cooling)
OFF
ON (cooling)
Status message
Hightemp.: 1
Pump-warn.: 1.2 (when
pump is connected
and configured
accordingly, see
Section 16.2)
OK
OK
www.mc-techgroup.com