12.4.3 CONNECTION OF HEATED LINE FOR PORTABLE VERSION (PART NO. 01G6190)
If, for the portable version in a carrying case, the option temperature controller for heated line (Part No.
01G6190) is selected, on the rear of the housing of the CSS-VC.. there is a 7-pin connector for
connection of heated lines type PSP and PSP 4M and PSP 4M-W. These have at the point of connection
and termination a 7-pin connector and a 7-pin socket for connection to the portable gas conditioning unit
CSS-VC.. and the portable gas sampling probe type PSP4000H. This is the power supply to the heated
line and gas sampling probe using the CSS-VC...
3 Mains
probe
2 Mains N
1 Mains L
Figure 10
Heated line connection for portable version to 10 A
12.4.4 CONNECTING THE OPTIONAL EXPANSION MODULES
The connection of expansion modules to the CSS-VC.. is via the communication module with the
appropriate cable to the RJ-45 socket in the connection panel (see Figure 3).
The electrical connection of the individual expansion modules is described in Section 15.
Notice
13
COMMISSIONING
Before commissioning, the plant and process-specific safety measures must to be observed.
Before switching on the power supply, it should be checked again that the operating voltage (see
nameplate) and the power supply voltage correspond!
Warning
www.mc-techgroup.com
The RJ45 socket is only suitable for the connection of the communication
module. A network connection is not possible with it!
Before comissioning exclude that for one channel as well internal as
external a PT100 is connected.
CSS-VC-1.6 | 1.01.03
4 Mains
probe
5 PT100
6 PT100
25