Temperature Alarms From Cooler And Temperature Control Module; Flow Disturbance And Liquid Alarms - M&C CSS Series Instruction Manual

Gas conditioning unit, 19“, wall mount or portable in carrying case with software version 1.6
Hide thumbs Also See for CSS Series:
Table of Contents

Advertisement

17.2
TEMPERATURE ALARMS FROM COOLER AND TEMPERATURE CONTROL
MODULE
Error message
Lowtemp.: 1
Hightemp: 1
T> Max: 1
T <min:1
Lowtemp.: 2, 3
Hightemp.: 2, 3
T> Max.: 2, 3
T <Min:2, 3
For channels 2 and 3 appears when the control temperature is fallen below (+10° C [+50 °F] alarm
hysteresis) of the following screen:
This prevents the heated line or the heated probe from overheating due to a defect! The triggered alarm
(red LED) must be acknowledged by pressing any key. If the error is not corrected, the error screen
comes back after 60 s.
17.3

FLOW DISTURBANCE AND LIQUID ALARMS

Error message
FA-Warn: 1, 2
LA-Alarm: 1, 2
LA-Break: 1, 2
In case of liquid alarm and activated LA-Reset (see 16.7) the following screen occurs :
www.mc-techgroup.com
Alarm
Warning
x
-
x
-
x
x
-
x
x
-
x
x
Alarm
Warning
-
x
x
-
x
-
CSS-VC-1.6 | 1.01.03
LED
Error
red
Low temperature cooler
red
High temperature cooler
red
Temperature sensor range exceeded
=> sensor defective
red
Temperature sensor measuring range
fallen below => sensor defective
yellow
Low temperature channel 2, 3
red
High temperature channel 2, 3
red
Temperature sensor range exceeded
=> sensor defective
red
Temperature sensor measuring range
fallen below => sensor defective
LED
Error
yellow
Flow disturbance channels 1 and/or 2
red
Liquid alarm channel 1and/or 2
red
Cable break liquid sensor channel 1 and/or 2
41