Other Fault And Alarm Messages; Maintenance - M&C CSS Series Instruction Manual

Gas conditioning unit, 19“, wall mount or portable in carrying case with software version 1.6
Hide thumbs Also See for CSS Series:
Table of Contents

Advertisement

This prevents liquid pumped along to the analyser and thus damaging ! The triggered alarm (red LED)
must be acknowledged by pressing any key. If the error is not corrected, the error screen comes back
after 60 s.
17.4

OTHER FAULT AND ALARM MESSAGES

Error message
Pump-Warn.: 1, 2
Maintenance needed!
Comm. Error
Flushing pressure not
present!
Thermal Error: 2, 3
18

MAINTENANCE

Before performing maintenance work, the system and process-specific safety measures must be
observed!
Warning
The maintenance intervals depend on the process conditions and therefore need to be determined for
specific applications.
42
Alarm Warning
x
x
-
x
x
-
-
x
x
-
Hazardous voltage. Pull the mains plug out before carrying out work on
the gas conditioning unit!
CSS-VC-1.6 | 1.01.03
LED
Error
red +
Pump 1 and/or 2 off due to an alarm
yello
(temperature, liquid)
w
yello
Maintenance interval expired
w
red
Error on connecting cable RJ45 between
CSS-VC.. and communication module
yello
Pressure for probe backflushing too low or
w
non-existent
red
Thermocouple set and defective or not
connected or PT100 connected
www.mc-techgroup.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Css-vc1Css-vc2

Table of Contents