This prevents liquid pumped along to the analyser and thus damaging ! The triggered alarm (red LED)
must be acknowledged by pressing any key. If the error is not corrected, the error screen comes back
after 60 s.
17.4
OTHER FAULT AND ALARM MESSAGES
Error message
Pump-Warn.: 1, 2
Maintenance needed!
Comm. Error
Flushing pressure not
present!
Thermal Error: 2, 3
18
MAINTENANCE
Before performing maintenance work, the system and process-specific safety measures must be
observed!
Warning
The maintenance intervals depend on the process conditions and therefore need to be determined for
specific applications.
42
Alarm Warning
x
x
-
x
x
-
-
x
x
-
Hazardous voltage. Pull the mains plug out before carrying out work on
the gas conditioning unit!
CSS-VC-1.6 | 1.01.03
LED
Error
red +
Pump 1 and/or 2 off due to an alarm
yello
(temperature, liquid)
w
yello
Maintenance interval expired
w
red
Error on connecting cable RJ45 between
CSS-VC.. and communication module
yello
Pressure for probe backflushing too low or
w
non-existent
red
Thermocouple set and defective or not
connected or PT100 connected
www.mc-techgroup.com