Setup For: Digital Ecg Unit - GE eBike Service Manual

Hide thumbs Also See for eBike:
Table of Contents

Advertisement

Transmission Protocol RS232 (PORT 1 and PORT 3): Digital EKG Unit
command from PC
(EKG unit)
Wxxx <CR>
S <CR>
s <CR>
B <CR>
D <CR>
O <CR>
U <CR>
F <CR>
I <CR>
- 76 -
Appendix A: Interfaces
feedback
Description
from eBike
sets nominal load value to xxx Watt
initiates BP measurement at rest, then starts the exercise
protocol with the initial load
initiates exercise protocol with the initial load, no BP
measurement at rest
B xxx <CR>
requests current nominal load
returns current load in W
n xxx <CR>
requests current RPM value
returns current RPM value
requests current systolic BP value
O xxx <CR>
returns current systolic BP value in mmHg
requests current diastolic BP value
U xxx <CR>
returns current diastolic BP value in mmHg
- - -
terminates exercise protocol
erxxxP10Vyyy
identifies the ergometer
<CR>
The string contains the ergometer model „xxx" and the software version.
Note: For compatibility with er900, the transferred string reads
„er900P10VV243".
eBike, eBike L, eBike EL
2018112-003 Rev J