Note: The Software Update menu displays the implemented software
version, FW 1.8
In addition, you can display a detailed overview of the software
versions of all installed modules.
Follow these steps to view the detailed software overview from the
Software Update menu.
PC Control Terminal: push bottom right softkey
PCPlus Control Terminal: push middle softkey on the right
Example of a detailed overview of the software modules
Software Module FW1.8
IFP: Rev: 5.00
AZP: Rev: 5.10
MRD: Rev: 5.20
NIBP: Rev: 1.58 (only on systems with NIBP option)
7. Error Codes
This function is currently not available.
Note: Error Code E:01 will be displayed after switching on the eBike,
when the cuff was moved during self-test. In this case the amplifier
cannot be initialized and the eBike has to be started again. E:01 will
not be recorded and will not appear here.
Note: If the error code E:01 appears again, even though the cuff was
not moved, the BP module is probably defective. To be able to run an
exercise test without the BP module, disable the blood pressure DIP
switch (see Figure C-6 DIP Switch and Jumper Settings).
Error Code E: 02 = MRD, offset or gain error
If a module version is detected that is not compatible with the installed software package, the boot
procedure will be aborted with one of the following error codes:
Wrong software in IFP board:
Wrong software in AZP board: the boot procedure is aborted and the error code E:91 displays.
Wrong software in NIBP board: the boot procedure is aborted and the error code E:92 displays.
Wrong software in MRD board: the boot procedure is aborted and the error code E:93 displays.
If the NIBP software does not works correctly during the regular use of the ergometer, the error code
E:10 displays.
If the MRD software does not work correctly during the regular use of the ergometer, the error code
E:20 displays.
Note: During the boot procedure, the software is checked in the above order. The procedure is
aborted, when the first incompatible software is detected. Only the first detected mismatch is
indicated, i.e., the remaining modules are not checked.
Remedy: Install software FW1.8 again as described in section 6.
- 62 -
D: Service Menu
the boot procedure is aborted and the error code E:90 displays.
eBike, eBike L, eBike EL
2018112-003 Rev J