GE eBike Service Manual page 110

Hide thumbs Also See for eBike:
Table of Contents

Advertisement

safety standards:
protection class:
MDD classification:
Environment:
operation:
relative humidity
atmospheric pressure:
transport and storage:
relative humidity:
atmospheric pressure:
Optional: BP Module
measuring principle:
cuff connection:
measuring range:
systolic pressure:
diastolic pressure:
pulse:
accuracy of the (static)
pressure readout:
resolution of the readout:
initial inflation pressure
pump rate:
max. cuff pressure:
cuff deflation method:
calibration:
artefact rejection:
- 104 -
Appendix C: Technical Specifications
BF to IEC 60601-1
I to IEC 60-601
class IIa to 93/42 EEC
temperature: +10...+40 °C/50...104 °F
30 ... 75%, 90% no condensation
700 ... 1060 hPa
temperature: -40 ... +70 °C (-40 ... +158 °F)
10 ... 95%, no condensation
500 ... 1060 hPa
- indirect measurement to Riva-Rocci-Korotkov
- oscillometric
the resting BP readings of both methods are compared for plausibility
tubing with connector and jack plug (microphone)
40 to 300 mmHg
40 to 300 mmHg
35 to 230 P/min
+/-3 mmHg
+/- 1 mmHg
during the inflation phase, the inflation pressure automatically adapts to patient's
BP
between approx. 6 seconds (to 140 mmHg) and approx. 18 seconds (to 300
mmHg)
300 mmHg
pulse-dependent deflation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
calibration with external pressure meter
automatic artifact rejection and comparison of the resting BP readings of both
methods for plausibility
eBike, eBike L, eBike EL
2018112-003 Rev J