4. Take off right and left pedals/pedal shoes.
a. Remove protective cover from pedals
b. Using the extractor (P/N 2005737-001), unscrew the
screw in the axle and remove (17-mm fork wrench
required)
c. Screw the extractor at least half way into the pedal
axle.
Turn the half of the extractor that points away from
the axle to the right until the crank can be removed.
5. Remove Casing
a. Loosen all casing screws - 3 on each side, see arrow
in illustration.
Important: DO NOT REMOVE the casing screws -
see red arrows - from the clamping rail.
Use two large screw drivers or two metal angles (side
length approx. 1.5 cm x 15 cm) to pull the casing
towards the outside at the pedal cutouts. Then pull
the casing first towards the front over the saddle motor
and then approx. 10 to 15 cm upward so that the
required cable and tubing connections can be
disconnected..
b. Disconnect the following cables:
-
Ground clamp at the bottom of the swivel section (see red arrow 1 in illustration at left)
-
Ground cable (see red arrows in illustration at right) (secured with adjustable cable tie)
-
eBike EL basic with blood pressure only: blood pressure tubing + microphone cable
-
eBike EL external only: disconnect cable for speed indication (RMP)
c. Carefully take off the casing.
6. Install Casing
Reverse the above steps to reinstall. Be careful to route the ground cable, blood pressure tubing +
microphone cable (option) and/or the cable for the RPM indication (option) correctly. When reinstalling
the casing, check that connection cable and air hose are not squeezed.