PCB Interface (item a.) and the pneumatic unit of the BP module (item b.) can be removed without
disassembling the device.
a. Remove PCB Interface
1. Disconnect device from mains!
2. Swivel handlebar to the front and tilt the ergometer 90°.
The ergometer rests on the handlebar and the base assembly
is accessible.
3. Disconnect all cables from PCB Interface (power cord and data cables).
4 Unscrew the 4 Philips screws and pull out PCB Interface.
5. If necessary, disconnect the cables from PCB Interface.
On the PCB you will find designations for the different connectors,
e.g. Messkopf X150.
b. Remove Pneumatic Unit of the Blood Pressure Module
1. Disconnect device from mains!
2. Remove PCB Interface, see item a.
3. Disconnect NIBP X201 cable from PCB Interface.
4. Unscrew the 4 Philips screws and remove the pneumatic unit.
5. Disconnect NIBP tubing.
c. Remove Top Cover
Two plastic tabs hold the top cover in place in the housing. By inserting
a tool such as a round pin with a max. diameter of 3 mm in a hole on
the left and on the right you can press these tabs towards the inside.
-
Using the round pins (max. diameter of 3 mm), simultaneously
press in the the plastic tabs on the left and right and
-
lift off the cover
2018112-003 Rev J
B: Mechanical Design
eBike, eBike L, eBike EL
- 15 -