Service Menu (accessible with password only after ergometer has been
turned on)
* Turn on the ergometer and simultaneously push the NIBP and MENU
buttons during the ergometer self-test until the Service Code menu
appears, prompting you to enter the service code.
* Using the arrow keys, enter the service code 999 and confirm entry
with the softkey ENTER.
When you entered a wrong service code, the message !! Wrong Code
!! appears and the ergometer enters the normal mode. If this happens,
turn the ergometer off and on again.
The menu offers the following items:
1.
Baudrate:
2.
Load Calibration:
3.
Analog Comparison: calibration for analog load control
4.
ECG-Startpulse:
5.
NIBP Parameter:
6.
Software Update:
7.
Error Codes:
8.
PC Mode:
9.
Saddle Calibration:
10. NIBP Calibration:
11. NIBP Test:
1.Select baud rate:
available settings:
Note:
Always select a baud rate of 4800 baud.
2a.
Load calibration eBike, eBike L, eBike EL of PCBRPM Board
1:
With software version FW1.6 the calibration procedure was redesigned.
In principle the procedure is similar to the previous one, but now the
technician will get more information and also more adjustments are
possible.
*
Take the seat ergometer to an elevated position, e.g. between
two solid chairs.
*
Support the supine ergometer so that the load unit will be in a
horizontal position.
*
Display the service menu and select menu item „Load
Calibration".
The screen shown at right appears.
The offset voltage of the WSG measuring device is displayed.
This voltage should be within a range of 60 mV to 80 mV.
If the value is not within that range, it should be adjusted by
means of a potentiometer on the MRD board (see section E-2.3
under „RPM Board: Replacement and Adjustment Procedure").
- 50 -
D: Service Menu
select baud rate for PORT 1
use 8-kg weight and calibrate load
adjust remote start for EKG unit
adjust NIBP parameters
update product software
not implemented
adjust PC mode
saddle calibration (eBike Basic, Comfort
only)
calibrate NIBP system
test NIBP system
1200, 2400, 4800, 9600, 19200, 38400,
57600, 115200 baud
eBike, eBike L, eBike EL
Note: When using
PCBRPM Board
2, please go to
section 2b.
Load Calibration
OFFSET
50
0
DMS-Voltage xx mV
SET
2018112-003 Rev J