Transfer Tubing Connection

the transfer tubing is routed from the syringe pump to the outlet of the automatic loop
injection valve of the PLC Purification System. refer to Plumbing Connections on page 96.
the transfer tubing is ordered separately, according to the syringe size. refer to the table
below for part numbers.
Part number
499671112
499474103
499483602
to install the transfer tubing:
1. Connect one end with 1/4"-28 fitting to the transfer tubing side of the valve of the
Verity 4020 Syringe Pump (right-hand port), and finger-tighten the nut.
2. Connect the other end to the outlet of the shut-off valve, connected to the waste outlet
(port 2) of the automatic loop injection valve of the PLC Purification System and finger-
tighten the nut.

Rear Panel Connections

Diagram
the following diagram provides a general overview of the connections described in this
section.
C
VeRiTY® 4020
Single Syringe
pump
Figure 73
Electrical Diagram for the PLC Purification System with GX-241 Liquid Handler and VERITY® 4020 Single Syringe Pump
98
PLC with autoSamPLer ConfiGuration
Syringe Size
Description
5 mL
Transfer tubing 5.5 mL
10 mL
Transfer tubing 10.5 mL
25 mL
Transfer tubing 30 mL
d
A
B
Gx-241
Liquid Handler
Transfer Tubing
Right Side
OR
e
G
|
PLC PurifiCation SyStemS
Left Side
F
pLC purification System