Install the Nebulizer and Nebulization Chamber
The nebulizer and the nebulization chamber are fragile. Unpack and handle them with care.
NOTICE
Ensure that each component is clean before installation.
to install the nebulizer and nebulization chamber:
1. remove the yellow cap from the evaporation tube.
2. remove the paraffin film from the nebulization chamber.
3. Position the nebulization chamber onto the stainless steel (SS) evaporation tube with the
siphon directed down. the chamber should be pushed as far as possible.
4. rest the nebulization chamber on its inclined support, and then tighten the small black nut
at the bottom.
5. Place the nebulizer on the top of the nebulization chamber, and then tighten the large
black nut with the gas inlet fitting directed down.
Figure 6
Removing the Evaporation Tube Cap
6. fill the siphon of the nebulization chamber with the mobile phase that will be used for the
separation. the liquid should fill the bend in the siphon, but not pool in the evaporation
tube. ensure that no liquid leak could affect the detector performance or create laboratory
pollution.
If using a volatile solvent (e.g., hexane or
NOTICE
dichloromethane), use water to fill the siphon.
evaporation
Tube Cap
PLC 2050/2250/2500 PurifiCation SyStemS
PLC 2050/2250/2500 PurifiCation SyStemS
inclined
Support
Figure 7
Positioning the Nebulization Chamber
Figure 8
Fill Area for Siphon Overflow
evaporation Tube
Small
Black nut
| uSer'S GuiDe
| uSer'S GuiDe
119