Split Flow Rates and Ratios

the following table provides results of tests performed on a PLC 2250 Purification System
(250 mL/min max.) using the 1/16" oD, 0.005" iD, 35 cm PeeK splitter tubing.
Split Flow Rates and Ratios with Acetonitrile (ACN)
eLution fLow rate (mL/min)
5
10
20
50
75
100
150
250
Split Flow Rates and Ratios with 90% Water - 10% Acetonitrile (ACN)
eLution fLow rate (mL/min)
5
10
20
50
75
100
150
250
SPLit fLow rate in eLSD (mL/min)
0.08
0.29
0.70
1.58
2.04
2.36
3.14
4.77
SPLit fLow rate in eLSD (mL/min)
0.12
0.24
0.47
0.80
1.01
1.21
1.61
2.35
PLC 2050/2250/2500 PurifiCation SyStemS
PLC 2050/2250/2500 PurifiCation SyStemS
SPLit ratio (%)
1.62
2.87
3.51
3.15
2.72
2.36
2.09
1.91
SPLit ratio (%)
2.32
2.44
2.33
1.60
1.34
1.21
1.07
0.94