Front Panel Connections

Connect the Siphon Overflow
1. Connect the supplied drain tube assembly
to the end of the siphon tube, using the
red, tapered hose connector with Ptfe
seal.
2. Lead the drain tube to waste.
NOTICE
120
120
PLC with inteGrateD eLSD ConfiGuration
PLC with inteGrateD eLSD ConfiGuration
Ensure that the waste container is directly below the siphon outlet and that there are no
loops or bends in the drain tubing.
Affix the drain tubing to the inlet of the waste container, so that the end of the tubing
never submerges into the liquid.
f the solvent you are using is not compatible with the drain tube, such as
tetrahydrofuran (THF), use a piece of PTFE tubing or any material compatible with your
solvent in its place. When using this type of tubing, which is generally more rigid, ensure
that it is safely installed, so that it will not damage the nebulization chamber (glassware).
Drain tube
Figure 9
Siphon Overflow Connections
|
|
PLC 2050/2250/2500 PurifiCation SyStemS
PLC 2050/2250/2500 PurifiCation SyStemS