Start Up

follow all prior instructions to make all rear panel and plumbing connections.
the Verity 1910 mS Detector has a safety interlock system to protect users from high
voltages. this system is operated by the door in the front cover.
when the door is closed, the safety interlock is off and high voltages can be applied. when
the door is fully closed it should be pushed gently forward until the catch engages with the
front cover. the catch is designed to prevent accidental activation of the safety interlock.
when the door is opened, the safety interlock is engaged and high voltages are disabled. the
door will click into position when the door reaches 90° and will remain fixed in this orientation
until pushed close.
to start the Verity 1910 mS Detector and the miDas:
1. Connect the power cords to grounded power outlets. the miDas turns on.
2. Switch on the monitor by pressing its 0/| button.
3. Close the mS detector door.
4. Switch on the mS detector by flipping the power switch on the rear panel (near power
socket 'P1').
5. Power on the mS detector by pressing the on/off power button (y) on the front panel.
● the green LeD status indicator on the front panel of the mS detector illuminates
(hardware interlock engaged).
● the green communication LeD on the front panel of the miDas illuminates.
● the monitor turns on and the masscape® Software starts.
6. in masscape Software:
a. Log in to Masscape Software as Superuser or above.
b. Select the Customise tab from the Masscape Software main screen.
c. Navigate to the Remote Options tab.
d. Ensure the options shown are selected or not
selected.
To perform calibration of the detector,
the VERITY 1910 MS Detector must be in
LOCAL mode. The complete procedure to
NOTE
perform calibration via Masscape Software
is detailed in the Masscape® Software
User's Guide "4500 MiD Software Guide
Version 3.1x".
7. Power on the PLC system and wait for
Gilson Glider Prep (GGP) Software to load.
if the PLC Purification System and the
Verity 1910 mS Detector are properly
configured and the communication established,
the masscape Software is in 'remote mode' (visible in masscape Software).
for more information about how to control a Verity 1910 mS Detector and
PLC Purification System using GGP Software, refer to the instructions document on the
documentation uSB shipped with the Verity 1910 mS Detector.
Figure 81
Customise tab
PLC 2050/2250/2500 PurifiCation SyStemS
111
| uSer'S GuiDe