Install the Chamber Shield

the chamber shield encloses the nebulization chamber and is held in place by two black
plastic holders in front. the chamber shield is designed to minimize glassware breakage and
prevent tubing deterioration. take care not to damage the splitter tubing, nebulizer, and
nebulization chamber, and then slide the shield diagonally downward.
to install the chamber shield slide the chamber shield into the exposed notches of the two
shield holders.
Figure 12
Installing the Chamber Shield

Operation

Start Up
make all rear and front panel connections before start up.
to use the eLS detector:
1. open the gas source.
2. unlock the regulating device of the gas regulator by lifting it to be able to adjust the gas
pressure supplied to the system.
3. turn the regulating device clockwise (+) to set the gas pressure to 2 bar (29 psi) on the
manometer.
4. once the gas pressure is adjusted, lock the regulating device by pressing it.
5. open the split way by switching the quarter-turn shut-off valve (small handle turned up),
that allows to divert a minor fraction of the purification flow to the eLS detector.
122
122
PLC with inteGrateD eLSD ConfiGuration
PLC with inteGrateD eLSD ConfiGuration
Shield
holders
|
|
PLC 2050/2250/2500 PurifiCation SyStemS
PLC 2050/2250/2500 PurifiCation SyStemS