Injection Valve Waste Outlet
NOTE
Used only for the automatic loop injection valve.
the injection valve waste outlet (Pn 21040169) tubing connects the waste outlet of the
injection valve to a waste reservoir. there is a fitting at one end and an SS ballast on the other
end. Connect the tubing to the shut-off valve outlet and finger-tighten the PeeK nut. the
shut-off valve is connected to the port 2 of the automatic loop injection valve.

Column Setup

a wide range of columns can be used with PLC Purification Systems, including normal or
reverse phase columns and those with an overall external diameter up to 40 mm. the device
used to install the column is a clamp with vinyl coating. the clamp is mounted on the black
holder on the left side of the system.
Clamp Holder Screws
Figure 31
Column Positioning
Figure 30
injection Valve Waste outlet
PLC 2050/2250/2500 PurifiCation SyStemS | uSer'S GuiDe
Waste
outlet
Shut-off
Valve
Wing
nut
41