When checking the over-temperature cut-out, be sure that the
temperature limits of the bath fluid are not exceeded. Exceeding
the temperature limits of the bath fluid could cause harm to the
operator, lab, and instrument.
12-2
If a hazardous material is spilt on or inside the equipment, the user is
responsible for taking the appropriate decontamination steps as outlined by
the national safety council with respect to the material. MSDS sheets
applicable to all fluids used in the baths should be kept in close proximity to
the instrument.
If the mains supply cord becomes damaged, replace it with a cord with the
appropriate gauge wire for the current of the bath. If there are any questions,
contact an Authorized Service Center for more information.
Before using any cleaning or decontamination method except those
recommended by Hart, users should check with an Authorized Service
Center to be sure that the proposed method will not damage the equipment.
If the instrument is used in a manner not in accordance with the equipment
design, the operation of the bath may be impaired or safety hazards may
arise.
The over-temperature cut-out should be checked every 6 months to see that
it is working properly. In order to check the user selected cut-out, follow the
controller directions (Section 8.2) for setting the cut-out. Both the manual and
the auto reset option of the cut-out should be checked. Set the bath
temperature higher than the cut-out. Check to see if the display flashes cut-
out and the temperature is decreasing.
 WARNING: