Fluke 7008 User Manual page 44

Calibration bath
Table of Contents

Advertisement

7008
Users Guide
The cut-out has two modes — automatic reset and manual reset. The mode
determines how the cut-out is reset which allows the bath to heat up again. When
in automatic mode, the cut-out will reset itself as soon as the temperature is
lowered below the cut-out set-point. With manual reset mode the cut-out must be
reset by the operator after the temperature falls below the set-point.
When the cut-out is active and the cut-out mode is set to manual ("reset") then
the display will flash "cut-out" until the user resets the cut-out. To access the
reset cut-out function press the "SET" button.
The display will indicate the reset function.
9-2
Access cut-out reset function

Advertisement

Table of Contents
loading

Table of Contents