7008
Users Guide
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trouble-
free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and cleaning
recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
•
•
•
•
•
•
3-2
temperature range: 5 to 50 °C (41 to 122 °F)
ambient relative humidity: maximum 80% for temperature < 31 °C,
decreasing linearly to 50 % at 40 °C
pressure: 75kPa - 106kPa
mains voltage within ±10 % of nominal
vibrations in the calibration environment should be minimized
altitude does not affect the performance or safety of the unit