Calibration bath
Table of Contents

Advertisement

D0 = −25.229
DG = 0.0028530
= 75.00 C
t
L
measured t =74.901 C
Compute errors,
= 24.869 − 25.00 °C = −0.131 °C
err
L
= 74.901 − 75.00 °C = −0.099 °C
err
H
Compute C
(
. 0
131
)(
75
0 .
(
'
D
0
Compute C
= '
D
G
,
0
25
.
229
))
(
. 0
099
75
0 .
25
0 .
,
G
(
. 0
099
)
(
. 0
131
)
 
75
0 .
25
0 .
Figure 10. Sample calibration computations
)(
25
0 .
(
25
.
229
))
+
=
1
. 0
0028530
. 0
0028548
 
Calibration Procedure
11.2 Measuring the Set-point Error
+
=
(
25
.
229
)
25
.
392
11
11-3