Fluke 7008 User Manual page 34

Calibration bath
Table of Contents

Advertisement

7008
Users Guide
1. This is the top half of the latch which attaches the lid to the bath. There are
2. The thermometer/liquid level is a hole in the lid for filling and removing of
3. The rectangular access hole may be used to insert the user's device to be
4. The circulation inlet - outlet is the access area or the inlet and outlet for the
5. The stirrer/pump motor is mounted on the bath tank lid. It drives the stirring
6. The probe hole is for the temperature controller probe. The probe must be in
7-6
two latches one on the front and one on the back.
liquid, checking liquid level and for mounting a thermometer. A hole may be
drilled in the rubber stopper supplied for inserting a thermometer into the
bath.
thermo stated. To obtain specified control levels this hole must be covered to
prevent thermal disturbance to the bath. Covers may be purchased
separately from Hart Scientific.
circulation pump.
propellers to provide mixing of the bath fluid. Proper mixing of the fluid is
important for good constant temperature stability and uniformity. The
stirrer/pump motor power cord plugs into the back at the power socket
labeled "STIRRER". The pump motor (on optional pump lid) is mounted on
the bath tank lid.
this hole for the bath to operate.