Table 4. Proportional Band — Fluid Table
Fluid
Temperature
30.0 °C
Water
60.0 °C
Water
35.0 °C
Eth-Gly 50%
60.0 °C
Eth-Gly 50%
100.0 °C
Eth-Gly 50%
35.0 °C
Oil
60.0 °C
Oil
100 °C
Oil
The proportional band adjustment may be accessed within the secondary menu.
Press "SET" and "EXIT" to enter the secondary menu and show the heater
power. Then press "SET" to access the proportional band.
+
12
Pct
Heater power in percent
Access proportional band
Pb=0.101C
Proportional band setting
To change the proportional band press "UP" or "DOWN".
Pb=0.060C
New proportional band setting
To accept the new setting and access the cut-out set-point press "SET". Pressing
"EXIT" will exit the secondary menu ignoring any changes just made to the
proportional band value.
Accept the new proportional band setting
Heater Setting
Low
Low
Low
Low
High
Low
Low
Low
Access heater power in secondary menu
Controller Operation
9.7
Proportional
Stability
Band
0.04 °C
±0.0004 °C
0.04 °C
±0.001 °C
0.05 °C
±0.0005 °C
0.05 °C
±0.001 °C
0.4 °C
±0.007 °C
0.1 °C
±0.003 °C
0.2 °C
±0.002 °C
0.2 °C
±0.003 °C
9
Proportional Band