Installation Procedure - Keysight Technologies X Series Installation Note

Signal analyzers cpu replacement procedure
Hide thumbs Also See for X Series:

Advertisement

Installation Procedure

1. Refer to
Figure
1. Locate and remove the disk drive tray assembly (2) from the instrument by
loosening the drive tray retaining screw (1) and pulling the assembly out by the handle on the
drive tray rear panel.
Set the drive aside in a safe location for replacement later in this procedure.
Figure 1
Removing the Disk Drive Tray Assembly
Installation Note N9020-90248
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on
electronic assemblies should be performed at a static-safe workstation. Refer to the
documentation that pertains to your instrument for information about static-safe workstations
and ordering static-safe accessories.
PC6 and PC7 CPU Replacement Kit
5