Olympus MVX10 Instructions Manual page 39

Research macro zoom system microscope
Table of Contents

Advertisement

2
Assembly
# Before assembly, remove dirt and dust from the parts and perform the operation cautiously.
Stage plate
³
Gliding surfaces
²
Stage plate
mounting hole
@
3
Operation
Fig. 29
Fig. 30
Notes
# Take care not to touch the gliding surfaces of the cup stage and mount with a hand. If they are contaminated,
clean with neutral detergent before use.
# The cup stage may move spontaneously if a load of 20 g or more is applied to its outer periphery.
# When the cup stage is tilted while a specimen with large height is placed on it, the image may be defocused. If
this happens, adjust focusing again.
Specimen holder
|
Specimen holder mounting hole
Cup stage
Mount
Applicable base
Place a specimen on the stage plate, hold the edge of the cut stage and
tilt it slowly. (Fig. 29)
}If the specimen slips off, fix it using the provided specimen holder.
}To fix a container such as a petri dish, fit the provided tube around the
specimen holder and hold the container with it. (Fig. 30)
@ Fit the cut stage mount into the stage plate
mounting hole of an applicable base.
² Place the cup stage on the mount.
Before placing, clean the gliding surfaces on
the cup stage and mount by wiping with a clean
cloth.
³ Attach the stage plate.
| Attach the specimen holder.
}Clean the gliding surfaces periodically.
MVX10
(Figs. 29 & 30)
35