10-1 Revolving Nosepiece MVX-2RE

}The revolving nosepiece accepts two MVPLAPO series objectives to enable observation using a wider variety of magni-
fications. The objectives can be switched quickly by simply rotating the revolving nosepiece horizontally.
1
External View
Nosepiece mount
2
Assembly
²
24

OPERATION OF OTHER MODULES

@
Fig. 20
Hand guard
Objective mounts
1. Remove the trinocular head and coaxial fluorescence illuminator from
the zoom microscope body.
2. Remove the zoom microscope body and focusing assembly from the
pillar, and place them upside down on a flat desk surface.
# Place a soft sheet made of rubber or similar material on the desk
surface.
3. Remove the objective from the zoom microscope body and then remove
the objective mount ² by loosening the four clamping screws @ using
the provided Allen wrench (for M3 screws). (Fig. 20)
Allen wrench
(for M3 screws)
Clamping screws
M3, 6 mm long (x 4)
(Fig. 20)