Olympus MVX10 Instructions Manual page 36

Research macro zoom system microscope
Table of Contents

Advertisement

2
External View & Specifications
3
Assembly
1. Attach the SZ2-FO vertical-movement stage inside the stage plate mounting hole of an applicable base by using the
provided clamping screws and the Allen wrench.
The stage can also be mounted so that the focus adjustment knob comes on the front side.
2. Loosen the clamping knob of the stage, attach the stage plate SZH-SG or SZH-SC and tighten the clamping knob again.
4
Operation
@
Push up.
Clamp.
32
Clamping screw holes
Stage
Load capacity: 1 kg
Clamping knob
Fig. 26
To increase.
To decrease.
Fig. 27
Removing the Stage Plate (Fig. 26)
Loosen the clamping knob @ and push up the stage plate from its
bottom side.
Adjusting the Rotation Tension of the Focus Adjustment Knob (Fig. 27)
}This adjustment is intended to facilitate the rotation of the knob while
preventing the vertical-movement stage from lowering spontaneously.
For best ease of use, it is recommended to adjust the rotation tension
slightly tighter than the level at which spontaneous lowering of the stage
occurs.
@
a. Holding the left and right focus adjustment knobs @ with both hands,
fix the left knob and rotate only the right knob to increase or decrease
the rotation tension.
b. If the rotation tension is too large, fine focusing will be unavailable
and the mechanism may eventually be damaged.
Focus adjustment knobs
Stroke: Approx. 21 mm
Rotation tension adjustable.

Advertisement

Table of Contents
loading

Table of Contents