Replacing The Probe Base - Icare TA022 Instruction Manual

Hide thumbs Also See for TA022:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
8
Error light
Probe base light
is flashing red
and Measure
light turns off
(Figure 34)
Repeat light is
flashing and
Probe base light
is flashing red
(Figure 35)
Service light is
flashing and
Probe base is
flashing red
(Figure 36)
The indicator lights on the back panel:
OFF (Figure 37)
1.
Lights off
INTIALIZE (Figure 38)
2.
LOAD (green icon)
3.
MEASURE (green icon)
4.
REPEAT (yellow icon)
5.
DONE (green icon)
6.
SERVICE (red icon)
7.
BATTERY (red icon)
LOAD PROBE (Figure 39)
8.
LOAD (green icon)
MEASURE (Figure 40)
9.
MEASURE (green icon)
REPEAT (Figure 41)
10.
REPEAT (yellow icon)
DONE (Figure 42)
11.
DONE (green icon)
SERVICE (Figure 43)
12.
SERVICE (red icon)
LOW BATTERY (Figure 44)
13.
BATTERY (red icon)
EMPTY BATTERY (Figure 45)
14.
BATTERY (red icon)

14. REPLACING THE PROBE BASE

Conditions during the use or transport of the tonometer can include lint,
dust, heat, moisture, hair, grease or other debris that can get to the probe
base and impair the probe movement. To resolve these conditions replace
the probe base at least every 6 months. Change the probe, replace the probe
base if the SERVICE light flashes.
Icare HOME Instruction Manual
Error
Reason
sound
Two long
Tonometer is
beeps
not correctly
aligned or
probe hit eyelid
or eyelash.
Three
Tonometer is
long
too far from
beeps
eye.
Five
Tonometer is
short
too close to
beeps
eye.
Two long
a) Too much
beeps
IOP deviation
during the
measurement,
because the
user did not
keep the
tonometer
stable.
b) Eye was not
recognized.
Two long
Damaged or
beeps
dirty probe or
probe base.
Action
1. Press the measurement
button again to clear the error
message.
2. Set the probe perpendicular
to the center of the cornea.
3. Adjust the forehead/cheek
supports for correct distance.
4. Repeat measurement.
1. Press the measurement
button again to clear the error
message.
2. Repeat the measurement.
3. Do not move the tonometer
during the measurements,
remove your hand or fingers
from the infrared transmitters
and sensor, move the patient's
hair away from his/her temple
side of the eye.
1. Change the probe and/or
probe base.
2. Contact provider of the
tonometer or a certified
service center.
ENGLISH