Error light
Probe base light
is flashing red
and Measure
light turns off
(Figure 35).
Repeat light is
flashing and
Probe base light
is flashing red
(Figure 36).
Service light is
flashing and
Probe base is
flashing red
(Figure 37).
OFF (Figure 38)
1.
Lights off
INTIALIZE (Figure 39)
2.
LOAD (green icon)
3.
MEASURE (green icon)
4.
REPEAT (yellow icon)
5.
DONE (green icon)
6.
SERVICE (red icon)
7.
BATTERY (red icon)
LOAD PROBE (Figure 40)
8.
LOAD (green icon)
MEASURE (Figure 41)
9.
MEASURE (green icon)
REPEAT (Figure 42)
10.
REPEAT (yellow icon)
DONE (Figure 43)
11.
DONE (green icon)
SERVICE (Figure 44)
12.
SERVICE (red icon)
LOW BATTERY (Figure 45)
13.
BATTERY (red icon)
EMPTY BATTERY (Figure 46)
14.
BATTERY (red icon)

13. REPLACING THE PROBE BASE

Replace the probe base every twelve months. Change the probe, replace or
clean the probe base if the Service light flashes.
Instructions for replacing the probe base:
• Turn off the tonometer.
• Unscrew the probe base collar and put it in a safe place.
• Remove the probe base by tilting the tonometer downwards and use your
fingers to pull the probe base out of the tonometer.
• Insert a new probe base into the tonometer.
• Screw the collar in, to lock the probe base.
Error
Reason
sound
Two
a) Probe is too
long
far from or too
beeps.
near the eye.
b) Probe
movement
was not
perpendicular
to the cornea.
Two
a) Too much
long
IOP deviation
error
during the
beeps.
measurement,
because the
user did not
keep the
tonometer
stable.
b) Eye was not
recognized.
Two
Incorrect or
long
dirty probe or
beeps.
probe base.
Action
Press the measurement
button again to clear the error
message.
a) Set the correct distance
4-8mm (5/32-5/16") between
the probe tip and the center of
the cornea.
b) Set the probe perpendicular
to the center of the cornea.
Press the measurement
button again to clear the error
message.
a) Repeat the measurement.
b) Do not move the tonometer
during the measurements,
remove your hand or fingers
from the infrared transmitters
and sensor, move the patient's
hair away from his/her temple
side of the eye.
Change the probe, clean or
change the probe base or
contact the seller to arrange
sending the device for service.