Electromagnetic Effects; Bladderscan ® Bvi 9400 Instrument Specifications - Verathon BladderScan BVI 9400 User Manual

Bladder volume instrument
Hide thumbs Also See for BladderScan BVI 9400:
Table of Contents

Advertisement

BladderScan
BVI 9400
®
Specification
Canadian Standards Association C22.2 No. 601.1-M90 (Master Contract No. 177198)
Medical Device Directive
Underwriters Laboratories, Inc.
*For details on Verathon
QuickPrint Help menu (select PRIVACY AGREEMENT).

Electromagnetic Effects

There are no restrictions on the use of BladderScan
characteristics. Both the emissions from BladderScan
to interference from other sources are within prescribed limits of all applicable standards at the
date of manufacture. The emissions test procedure that was used is specified in EN/ IEC55011:
1991 for Group 1, Class A equipment (per EN/IEC60601-1-2, 36.201.1.7).
BladderScan
environments, and in domestic environments under the jurisdiction of a health care professional.
An indication of adverse electromagnetic effects from BladderScan
electronic device would be a degradation of performance in the other device when the devices
are operated simultaneously. If such interference is suspected, separate the two devices as
much as possible, or discontinue simultaneous operation, if practical, and contact Verathon
BladderScan
sources, and has demonstrated immunity at a field strength of 3 V/m (per EN/IEC 60601-1- 2,
36.202.2.1). You do not need to take any other precautions regarding exposure in reasonably
foreseeable environmental conditions to magnetic fields, pressure, or variations in pressure,
acceleration, or thermal ignition sources.
BladderScan
BVI 9400 Instrument Specifications
®
For technical performance assessment in an acoustic laboratory, the Probe cannot be immersed
in water.
Item
Input
Output
Insulation
Transient Overvoltage
Weight
User's Manual
Standard
MDD 93/42/EEC Annex 1
UL 60601-1
Health Insurance Portability and Accountability Act (HIPAA)*
compliance with privacy rules, please refer to the information in the
®
BVI 9400 is suitable for use in industrial, scientific, and medical (ISM)
®
BVI 9400 will operate normally in the proximity of other potential interference
®
Specification
Lithium Ion Battery Pack. Verathon P/N 0400-0066.
No load to full load at rated voltage. Refer to unit label.
The power supply is Class I with basic insulation to each terminal.
Category II
5.2 lb (2.36 kg) (with battery)
Product Specifications
BVI 9400 due to its electromagnetic
®
and the susceptibility of this instrument
®
BVI 9400 on another
®
.
®
page 65