Test Procedure Overview - Benning ST 755 Operating Manual

Appliance tester
Table of Contents

Advertisement

13.2

Test procedure overview

CLASS I
RPE
(protective conductor resistance)
RInsu-1 (insulating resistance LN-PE)
RInsu-2 (insulating resistance sec.-PE)
RInsu-3 (insulating resistance LN-sec.)
RInsu-4 (insul. res. LN-access. parts without PE)
ILeak
(device leakage current)
PLeak
(patient leakage current)
Funct.
(functional test)
Cable
(continuity test)
Test failed
VDE 0751-1
Devices with PE,
Class I
Visual inspection
(standard, advanced,
customer-specific)
No
Visual inspection passed?
Yes
No
Test passed?
Yes
Test successful
Devices without PE,
Class II
CLASS II
RInsu-1 (insulating resistance LN-test body)
RInsu-2 (insulating resistance sec.-test body)
RInsu-3 (insulating resistance LN-sec.)
ILeak
(device leakage current)
PLeak
(patient leakage current)
Funct.
(functional test)
Cable
(continuity test)
Specification, selection
individual tests are
determined by the
qualified electrician
carrying out the test.
Individual tests:
and order of the
31