10 OPTIONAL FUNCTIONAL TEST OF 651/631-X

AMETEK Brookfield has a 30 ppb Hydrogen Sulfide Functional Test
Module (FTM), P/N Z2600 0930, used to determine if your instrument
is functioning correctly between recommended factory calibrations. It
allows you to have complete confidence in the sample results. This
test verifies proper instrument operation through the introduction of a
known concentration of Hydrogen Sulfide gas. Call Jerome Sales at
(800) 528-7411 or (602) 470-1414 for availability and pricing.
THIS VERIFICATION DOES NOT CALIBRATE THE INSTRUMENT.
If your application requires frequent verification of instrument function, this test
demonstrates the unit's operation. Recording FTM results in an instrument log provides
a quality control/quality assurance record of instrument function between regular
calibrations. If test results fall within the expected range, you may assume the
instrument is functioning correctly.

10.1 Test Procedure

1. Warm up the FTM indoors for six (6) hours to ensure a stable output. Refer to the
FTM operation manual, P/N 700-0095.
2. After warm-up, use a portable power supply to keep the FTM running while checking
the 651's. (The installation of a weatherproof electrical receptacle near the
monitoring system is a recommended alternative.)
3. Open the 651 door, attach the S shaped hanger (P/N 835-0024) to the 651, hang the
FTM on the hanger, and attach the FTM sample tube to the 651 intake at the
bottom.
4. Press [Select] to enter the menu system; the 651 displays "FTM CHK."
5. Press [Select] again; the 651 displays "CONFIRM?"
6. Press [Select] once more and the 651 will automatically take seven (7) samples.
7. Average the last (5) samples and the average should be within 26 to 34 ppb.
8. Disconnect the FTM from the 651 and reconnect any "bug plugs" or drip controls to
the intake tube.
9. Proceed to the next 651 (if applicable).