3.0 Functionality

3.1

Functional description

The probes can be used for temperatures in the
range :
• DLC-155 : -60°C to 155°C / -76°F to 311°F
• DLC-158 : -60°C to 155°C / -76°F to 311°F
• DLC-159 : -100°C to 155°C / -148°F to 311°F
• DLC-168 : -60°C to 165°C / -76°F to 329°F
• DLC-187 : -60°C to 180°C / -76°F to 356°F
• DLC-250 : 0°C to 250°C / 32°F to 482°F
• DLC-700 : 0°C to 700°C / 32°F to 1292°F
The JOFRA DLC-155, DLC-158, DLC-159, DLC-
187, DLC-250 and DLC-700 probes may be supplied
with a certificate for a limited temperature range.
The calibration temperature ranges for the DLC
sensors are stated on page 19.
127414
06
2021-10-29
Page 5