Advertisement

4. Enter the password and press the ENT key.
5. Move the * mark next to Self-check.
You can move the cursor using the
Keys.
6. Press the ENT key.
7. Press the START key. (Begins the Self-Check)
8. The Self-Check ends (If there are no errors in any of the valves or sensors, "No error" will be
displayed.)
MAINTENANCE FUNCTIONS
Maint1/2P
*Volume Measurement
I/Ocheck
SelfCheck
StabilityCheck
SoftVersion
Maint1/2P
Volume Measurement
I/Ocheck
*SelfCheck
StabilityCheck
SoftVersion
SelfCheck
1 Cal Port Open
2 ValveClose
3 Press Start Ke y
SelfCheck
0%
SelfCheck
No error
SELF-CHECK FUNCTION
100%
81