POWER ON PROCEDURE Connection to External Equipment
REMOTE JUDGMENT CHART
■
・The timing and explanation for the judgment results that are indicated on the main display and/or output from the
REMOTE Terminal.
Status
Setting
Contents
LEAD.P =
Work internal pressure is
Lead
more than ±10% out of
pressure
the Lead setting value.
value
TEST.P =
Work internal pressure is
Test
more than ±10% out of
pressure
the Balance 1 value.
value
Differential
sensor indicates less
than –FS.
Differential
sensor indicates less
than –FS.
Measurement result
-NG
Measurement
-NG
<
result
<+NG
+NG<
Measurement
+2NG
+NG
≦
result
<+2NG
Measurement
+2NG ≦
result
Differential
sensor indicates more
than +FS.
Differential
sensor indicates less
than –FS.
Measurement result
-NG
Measurement
+NG=
-NG
<
result
<+NG・+2NG
+2NG
Measurement
+NG・+2NG≦
result
Differential
sensor indicates more
than +FS.
44
Judgment
Result
Judging timing
Pressure 1 Pressure 2 Balance 1 Balance 2
pressure
pressure
≦
pressure
pressure
≦
pressure
F P.NG
+00.0
Hold
△:Judging point
←→ :Judging range
Detection
(*1)
No update
kPa
Group No 00
Judgment
result shown
on LCD
Terminal
Exhaust
display
P.NG(*3)
P.NG(*3)
B2.+NG
B2.-NG
D-NG
-NG
OK
+NG
+2NG
D+NG
D-NG
-NG
OK
+2NG
D+NG
REMOTE
P.NG
P.NG
B.NG
-NG
-NG
OK
+NG
+2NG
+2NG
-NG
-NG
OK
+2NG
+2NG