BASIC OPERATIONS
2. Out Data 2 (Detection Process (DET、MDET) data, test result transmission)
A differential pressure value (Pa value) is transmitted every 0.1 sec during the DET phase.
The test result data is transmitted during the EXH process, after the testing is complete.
(The format of the text transmitted for the test result is the same as with Out Data 1)
Format of transmitted text (Detection Process (DET, MDET) data)
Item
Pressure Value(Pa)
Byte No.
Data side
3. Out Data 3 (Transmits data and test results from the beginning of the CHG1 phase through the
DET (MDET) phase.
A work pressure value (kPa value) is transmitted every 0.1 sec during the CHG1, CHG2, and BAL 1
processes,
and a differential pressure value (Pa value) is transmitted every 0.1 sec during the BAL2 and DET
(MDET) processes.
(The format of the text transmitted for the test result is the same as with Out Data 1)
Transmitted text format
(Work Pressure value)
Item
Byte No.
Data Ex.
○ Test value transmission
The transmitted test value contents include all the setting items for the currently selected
group.(see page 51~53)
Transmitted text format
Item
Byte No.
Data side
Balance2(sec)
WorkVolume(ml )
Remote Stop
The transmission process is shown below.
1. Press the MAINT key at the initial screen.
2. Input the password and press the ENT key. Select Data Load on page 2/2 using the
3. Press the ENT key.
4. Press the START key.
66
RS232C OPERATIONS
Differential
6
+00000
Work Pressure
Value(kPa)
6
+00.0
-
Delay(sec)
CR Charge1(sec) CR Charge2(sec) CR
6
1
000.0s
0DH
CR
6
1
001.0s
0DH
CR
7
1
1.000E2
0DH
-NG(ml/min)
CR +FNG(ml/min) CR -FNG(ml/min) CR UnitChange CR
6
1
-1.000
0DH
Lead.P(kPa)
CR
6
1
+000.
0DH
-
CR
2
1
0
0DH
-
ETX
Space
1
1
03H
20H
CR
1
0DH
Transmitted text format
(Differential Pressure value)
Item
CR
Byte No.
1
Data Ex.
0DH
6
1
000.0s
0DH
Detect(sec)
CR
Exhaust(sec)
6
1
002.0
0DH
CR +NG(ml/min) CR +2NG(ml/min) CR
Master
volume(ml)
12
1
1.000E2
0DH
-----
6
1
+1.000
0DH
Test.P(kPa)
CR
Fitting Set
6
1
+000.
0DH
-
UnitChange
CR
6
1
ml
kPa
0DH
-
CR
1
0DH
Differential Pressure
Value(Pa)
6
+00000
Balance1(sec)
6
1
002.0s
0DH
CR
FDetect(sec)
6
1
001.0s
0DH
6
1
+1.000
0DH
6
1
-1.000
0DH
CR
Out Data
1
1
0
0DH
Cylinder
CR
2
1
00
0DH
-
CR
1
0DH
CR
6
1
001.0s
0DH
CR
6
1
030.0s
0DH
6
1
+2.000
0DH
6
1
ml
kPa
0DH
-
CR
1
1
0
0DH
key.