Fukuda FL-601 Series Operational Manual page 58

Multipurose air leak tester

Advertisement

■ Master Volume Entry
This is an explanation of how to modify the Master Volume (M. Volume) setting items, which differs from the above
description.
1.
Use the
Pass 0000
Detect
Exhaust
F Detect
W.Volume
M.Volume0
2.
Press the ENT key.
( S elect the single-figure next to M. Volume using the cursor)
Pass 0000
Detect
Exhaust
F Detect
W.Volume
M.Volume0
Use the
3.
0:A fixed value of 3.77ml( this occurs when the master valve is closed) will be established as the Master
Volume.
1:A fixed value of 29.97 ml (this occurs when the accessory master tank is connected) will be established
as the volume.
2 : A fixed value of 53.33 ml (this occurs when the optional master tank is connected) will be established as
the volume.
3: The Master internal Volume is established manually. Press the
chosen.
The cursor will move to the following position.
Pass 0000
Detect
Exhaust
F Detect
W.Volume
M.Volume3
The Master Volume will be set up in the same way as W.Volume, by using the
4. Press the ENT
key.
( S elect the single-figure right next to M. Volume)
Pass 0000
Detect
Exhaust
F Detect
W.Volume
M.Volume3
keys to move the cursor to M. Volume.
2/5p
Gp 00
001.0 sec
001.0 sec
008.0 sec
5.000E2 ml
2.997E1ml
2/5p
Gp 00
001.0 sec
001.0 sec
008.0 sec
5.000E2 ml
1.000E1ml
keys to select a number from 0 to 3.
2/5p
Gp 00
001.0 sec
001.0 sec
008.0 sec
5.000E2 ml
1.000E1 ml
2/5p
Gp 00
001.0 sec
001.0 sec
008.0 sec
5.000E2 ml
1.000E1
ml
BASIC OPERATIONS SETTING PROCEDURES
See Operations 1 to 5 in the previous description for
help in navigating to this screen.
key once M.Volume 3 has been
keys.