■RS-232C Output
Data is output from the RS-232C connector on the front panel.
Types of data transmitted:
・ Setting data saved in the tester.
・ Test results (including the fitting test results) and test data.
○ Specifications for the RS-232C communications:
・Speed
・Start Bit
・Stop Bit
・Data length
・Parity
・Error Check
・Text Encoding
・Communication mode
・Transmission Process
・Connector
○ Connector Chart and Wiring
RS232C Wiring
Tester
Signal Name
Number
D-Sub 9
DCD
TxD
RxD
DSR
SG
DTR
CTS
RTS
RI
* If the pin configuration of your sequencer, etc. is different from that of the
computer, ensure that the appropriate signals are connected.
The leak tester's RS-232C connector is a D-sub 9 Pin female.
○ Data Output Mode Setting
There are 4 modes for outputting data from the RS232C connector, depending on the setting.
You must set the data output (Data Out) mode according to your needs.
(See page 52 for more details.)
○ Test Value Transmission
1. Mode 1 Test value transmission (If OutData is set to 1)
The test result data is transmitted during the EHX process, after the test is completed.
Transmission Text Type
Item
Group Number
Byte No.
Data Ex.
+0.000
Comma
* The [_] mark signifies a space, ETX is equal to 03h, CR is equal to a Carriage Return(0dh),
and BCC is horizontal parity.
Pin
Connection
1
2
3
4
5
6
7
8
9
Comma
2
1
00
,
Test Result
Comma
13
1
ml/min
,
-
b Constant
1
10
,
+0072
Pa
-
--
BASIC OPERATIONS
19200 bps
1 bit
2 bit
8 bit
None
BCC (Horizontal parity)
ASCII
RTS・CTS Hand Shake
Non-procedural, asynchronous
9Pin D-SUB
Pin
Number
D-Sub 9
Judg
Comma
4
1
+2NG
,
Work Internal Pres.
Comma
9
1
+10.0
kPa
,
-
Comma
ETX
1
1
,
03H
RS-232C OUTPUT
Computer
Pin
Signal
Number
Name
D-Sub 25
1
8
DCD (CD)
2
3
RxD (RD)
3
2
TxD (SD)
4
20
DTR (ER)
5
7
SG (GND)
6
6
DSR (DR)
7
4
RTS (RS)
8
5
CTS (CS)
9
22
RI (CI)
Data Type
Comma
5
1
Mdata
,
C Invariable
10
-7.8441E-3
BCC
CR
1
1
0DH
(JIS)
65
Need help?