Constant Temperature Mode Operation

1. Plug the LDC-3706 Instrument into an appropriate AC power source supplying the correct
mains voltage and frequency for the instrument (refer to the rear panel for the correct ratings)
2. Turn on the LDC-3706; the TEC output will be disabled at power up and the unit will
automatically configure its parameters to the state which existed when the power was last
shut off
3. Press the SELECT soft button in the TEC MODE section of the front panel until the TEMP
LED is selected
4. Press the TEC soft button in the PARAMETER section of the front panel to adjust the
applicable limits, sensor type, sensor calibration constants, and external fan control
5. Adjust the temperature setpoint by enabling the ADJUST KNOB for TEC parameter
adjustment by pressing the TEC button in the ADJUST section of the front panel (upper
right). Use the ADJUST KNOB to change the setpoint temperature to the desired value by
rotating clockwise to increase the parameter and counter-clockwise to decrease. The new
parameter is now stored in non-volatile memory
6. The user can display measured voltage, current, or sensor value or the temperature setpoint
on the TEC side display (LEFT) and cycle through the available display parameters by
pressing the LEFT (left) and RIGHT (left) display buttons
7. Enable the TEC output of the LDC-3706 by pressing the ON soft button (left) in the TEC
MODE section of the front panel. This button has a toggling action so pressing this button
once more will disable the TEC output.