Set up Auto Decontamination routine

Select "Auto Decontamination" and click on
to set up a daily decontamination routine.
The following parameters can be selected
Start time
Weekdays
Functions
Select "Start time" and press
value starts flashing.
Pressing the
or
to ON. To activate or deactivate the daily
decontamination routine.
Press
to confirm and set "hours". Press
to confirm and set "minutes". Press
to confirm and set decontamination time between
30 and 5 minutes in steps of 5 minutes. Press
to confirm and store the settings. Press
to go back to the decontamination routine
parameters.
Select "Weekday" and press
Weekdays the auto decontamination will be
activated.
First character is flashing (M = Monday), pressing
or
button will modify the character from
active (filled box) to inactive (empty box). Press
to confirm and select the next day. Press
to confirm and select the next day after
selecting the last value (S = Sunday)
store the settings. Press
decontamination routine parameters.
VWR Collection PCR Workstation
. The first
button will modify from OFF
. To select the
to go back to the
will