INTRODUCTION

The VWR PCR Workstation is designed as an ideal environment for the manipulation of DNA and
RNA, especially for the set-up of PCR assays. Contamination can lead to false or misleading results
which costs time and money. The VWR PCR Workstation minimises the risk of contamination. It
provides a 'separate room', e.g. for the set-up of PCR reactions.
The high intensity surface UV tubes inactivate DNA as a source of contamination between
experiments. In addition, a UV Air Recirculator is integrated into the workstation system which
reduces airborne contaminants during experiments. UV light is effectively blocked by solid
polycarbonate screens to ensure maximum protection for the user.
The surface of the workstation is constructed of stainless steel, which is very robust and can easily be
cleaned plus has an antimicrobial effect.
The VWR PCR Workstation offers a controlled environment for PCR and RNA applications that
protects your samples and helps to achieve optimal results.

GENERAL WARNING

The PCR Workstation has been designed for laboratory use and it should not be used at temperature
below 15 °C. If the instrument is at a temperature below 10 °C place it in a warmer room and wait at
least two hours before using it. Before plugging the instrument into a power supply, check that the
voltage and frequency on the label matches the power supply specifications. Before any operation
ensure that the ventilation grids are not obstructed and the door is properly closed.

SAFETY INSTRUCTIONS

The PCR Workstation is designed with function, reliability, and safety in mind.
Before using the PCR Workstation please read this entire manual and strictly follow the safety
instructions!
Caution: The unit includes shortwave UV, which is a powerful source of UV radiation that will cause
damage to unprotected eyes and skin if directly exposed to the UV lamps. Before operating any unit,
ensure that the PCR Workstation is not damaged, and that instructions for use of this equipment are
followed. A safety shut-off switch automatically turns the UV light off when the front panel is open,
protecting users from UV exposure. The polycarbonate panels are UV-blocking.
Intended use of the PCR Workstation: Working environment for protection of PCR reactions /
reaction set ups and genetic substances against pollution and contamination.
Safety and protection of the user cannot be ensured if the PCR Workstation is used in any manner
that is not specified in these instructions. Special notice must be taken of the following:
VWR Collection PCR Workstation
Caution: Dangerous voltage
Ensure that the voltage requirements of the PCR Workstation exactly match your local
AC power supply.
Only use the original AC cable supplied with the PCR Workstation.
Caution: Liquids
Ensure that no liquids can enter the device. Do not place liquids on top of the PCR
Workstation.