Servicing schedule

This chapter describes the maintenance work to be performed on a yearly basis.
A checklist can be found in the attachment A-1.

Maintenance routine without dismantling the centrifuge

Electrical installation and safety
Switch OFF the centrifuge and disconnect the unit from power, check voltage supply and mains
fusing (16 (15) Amps, slow blow characteristic).
Check condition of plug and wall socket - (let) replace defective parts.
Check cord condition and fixing / connection - replace or refit it.
Check condition of instrument socket and replace it in case of bad contacts.
Location and mechanical installation
Check the base (ground, table, trolley with lockable wheels etc.) for resonance-free and stable
conditions.
Check for a well ventilated place and sufficient distances to walls or adjacent equipment, without
exposition to direct sunlight.
Check the leveling of the centrifuge drive with help of a spirit level.
Lid latch mechanism and safety circuit
Connect the centrifuge to power and switch ON.
Check for correct lid closing and opening - if in disorder, readjust lid, hinges and/or lid latch.
Check the rubber gasket for correct sealing and replace, if damaged.
Check gas lid stay for correct function.
Check the electronic safety circuit: start the centrifuge. Let it shortly run and press stop. The lid
must not be unlocked by the microprocessor as long as the speed is more than 60 rpm - if safety
circuit is out of function, replace main board.
Cleanliness of rotor chamber and motor cover
Open the lid and remove the rotor by pressing the green AutoLock button in the center of the
rotor.
Clean the spin chamber with a dry and absorbent cloth (remove all dust and moisture - see also
page 2-18 for cleaning).
Check the cleanliness of the motor cover and take special care of the slot between the bowl and
the motor cover: penetrating fluids can damage the motor and/or electronics and foreign objects
may block the air flow.
Condition and sealing of rotors and accessories
Check the condition of rotors and accessory parts (especially all supporting or stressed
partitions): the rotor and/or accessory parts must not be used any longer, if there are visible
traces of mechanical damage or corrosion.
Check the condition of rotor and/or accessory sealing and replace them in case of malfunction.