14.0 Notes on EMC

i
■ Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc.
to following EMC notes.
■ Portable and mobile HF communication facilities can influence medical electrical equipment.
■ The use of other accessories, other converters and cables than stated may lead to an increased emission
or a reduced interference immunity of the equipment or system.
14.1 Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS
A / C 161 Battery is intended for use in the electromagnetic environment specified below. The customer
®
or user of the ATMOS
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonics IEC 61000-3-2
Flicker
IEC 61000-3-3
24
A / C 161 Battery should ensure that it is used in such an environment.
®
Compliance
Group 1
Class B
Inapplicable
Inapplicable
Electromagnetic Environment - Guidance
The ATMOS
A / C 161 Battery uses RF energy
®
only for its internal function. Therefore, its RF emis-
sions are very low and are not likely to cause any
interference in nearby electronic equipment.
The ATMOS
A / C 161 Battery is suitable for use
®
in all establishments, including domestic, and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.