Installation; Environmental Conditions; Handling And Transportation; Unpacking - NUVE NB 5 User Manual

Water baths
Table of Contents

Advertisement

NB 5 Fuse (2x4A)

5. INSTALLATION

5.1 ENVIRONMENTAL CONDITIONS

The instrument is designed to operate safely under the following conditions:
 Indoor use only
 Ambient temperature: 5°C to 40°C
 Maximum relative humidity for temperature up to 31°C: 80%
 Maximum altitude: 2000 m
 Temperature for maximum performance: 15°C / 25°C

5.2 HANDLING AND TRANSPORTATION

All handling and transportation must be carried out by using proper equipment and experienced
staff. The instrument must be supported underneath and never be turned upside down.

5.3 UNPACKING

Remove the cardboard box packing and the second nylon wrapping around the instrument.
Ensure that no damage has occurred during transportation. The below mentioned are provided
with the instrument, please check them;
 1 ea. user's manual and warranty
 1 ea. electrical cable
 1 ea. perforated grid

5.4 MAINS SUPPLY

The instrument requires 230 V, 50 / 60 Hz.
Please make sure that the supplied mains matches the required power ratings which are written
on the name of plate of the instrument located at the back of the instrument.
Always plug-in the instrument to correctly grounded sockets.
A supply fitted with a circuit breaker should be used for protection against indirect
contact in case of isolation fault.

5.5 POSITIONING

 Check that the positioning is suitable for the users.
 Check that the instrument is stable on its four pedestals.
 Check that the user will be able to follow up the operation even when he deals with something
else.
NB 9 Fuse (2x5A)
NB 20 Fuse
Earthed Wall
(2x10A)
8
Sockets