9.1 ERROR CODES EXPLANATIONS

Er1
The temperature sensor endings are broken. The error code flashes on the temperature display
and an audible alarm sounds.
Er2
An electronic failure occurs in the microprocessor. The error code flashes on the temperature
display and an audible alarm sounds.
Er3
The temperature sensor measures a temperature higher than 147°C.The error code is shown on
the temperature display and an audible alarm sounds.
Er4
The temperature sensor endings are connected in reverse. The error code flashes on the
temperature display and an audible alarm sounds.
EoF
This error code appears if any probable power cut causes when the sterilization phase "EoF"
flashes and the audible alarm sounds on the temperature display.
IN CASE OF ANY ERROR, THE PROGRAM İS STOPPED AUTOMATİCALLY AND
IMMEDIATELY.
If an error occurs, please contact with an authorized Nüve agent to seek technical
help.