4.1 LIFTING AND TRANSPORT

All lifting and transport must be carried out by using proper handling equipment. The instrument
must be supported from underneath and never turned over.

4.2 UNPACKING

Remove the packing cardboard box and the second nylon packing around the centrifuge. The
below written are provided with the instrument, please check them;
1 ea. User's manual
1 ea. Warranty certification
1 ea. Power cable
1 ea. Manual lid opening tool

4.3 POSITIONING

Check that no damage has occurred during transport.
Check that the positioning is suitable for users.
Lift the centrifuge underneath and carry it to its place carefully. Check that the centrifuge is
stable on its four pads.
Leave sufficient free space on each side of the centrifuge.
Make sure that the centrifuge does not do harm to nearby equipment.
Nobody and dangerous material should not be inside of the security circle that
diameter is 300mm, according to IEC 61010-2-020.
4.4

ENVIRONMENTAL CONDITIONS

The centrifuges are designed to operate safely under the following conditions:
Indoor use only
Ambient temperature: 5
Maximum relative humidity of 80% for temperature up to 22 °C.
Maximum altitude: 2000 m.
Temperature for maximum performance: 15 °C / 25 °C.
4.5

MAINS SUPPLY

The centrifuge requires 230 V, 50 / 60 Hz.
Please make sure that the supplied mains match the required power ratings. If no, provide an
extra line to support.
Always plug the sterilizer/ovens to properly earthed sockets.
A supply fitted with a circuit breaker should be used for protection against
indirect contact in case of an insulation fault.

SECTION 4

INSTALLATION PROSEDURE

°
°
C to 40
C.
8