Emc-Compliant Cables And Accessories - GE MAC 800 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 800:
Table of Contents

Advertisement

Rated Maximum Output Power
of Transmitter in Watts
0.01
0.1
1
10
100
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

EMC-Compliant Cables and Accessories

2031504-159D
according to the maximum output power of the communications
equipment.
Separation Distance in Meters (m) According to Frequency of Transmitter
150 kHz to 80 MHz
150 kHz to 80 MHz
outside ISM bands
in ISM bands
d = 1.2
d = 1.2
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (w)
according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
WARNING
USE OF CORRECT ACCESSORIES — The use of
accessories, transducers and cables other than those
specified may result in increased emissions or decreased
immunity performance of the equipment or system.
The following table lists cables, transducers, and other applicable
accessories which GE Healthcare claims are EMC compliant.
NOTE
Any supplied accessories that do not affect EMC compliance are not
included.
MAC™ 800
Electromagnetic Compatibility
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2
d = 2.3
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
B-5