Q: What is the capacity of the battery?
A: We recommend that the MAC 800 be connected to AC power through
a wall outlet whenever it is not in use. However, if operating the device
without AC power, be aware that a fully-charged battery is capable of
printing approximately 1000 single page reports or 2 hours of continuous
operation (without printing).
Q: I need to provide the MAC address of the device to the network
administrator to enable the LAN communication option. How do I obtain
the MAC address?
A: Follow these steps to obtain the MAC address:
1. Open the Diagnostic Tests window as described in
System Diagnostics Function"
2. Move the focus to the Service Report button and press the Enter key
to select.
3. Find the MAC address on the printed service report.
Q: How do I calibrate the MAC 800 system?
A: When it becomes necessary, you can calibrate the MAC 800 system
using the following method:
1. Using a standardizing waveform generator, produce a 1.00 ± 0.01
-mV pulse signal with a rise-time no greater than 5ms and a width
no greater than 100ms.
2. Connect the pulse signal to all available channels and set the gain to
10 mm/mV.
3. Verify the display pulses have an amplitude within ± 5% of the
amplitude obtained when the 1.00 ± 0.01 -mV signal is applied.
4. Repeat the test for all fixed gain settings to verify the
standardization pulse correctly reflects the gain setting.
NOTE
The error must be less than ±5% of the expected value or 0.5mm,
whichever is greater.
5. Verify that the standardization signal appears on all channels.
MAC™ 800
"Accessing the
on page 3-6.
Troubleshooting
3-21

Hide quick links: