Functional Checkout - GE MAC 800 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 800:
Table of Contents

Advertisement

Maintenance

Functional Checkout

4-24
4. Reassemble the fuse holder into AC inlet.
The checkout procedures apply to all MAC 800 systems.
NOTE
The FRU checkout procedure for any listed FRU also applies to its
internal PCBs and components.
Perform the applicable product or product configuration dependant
procedures when an asterisk (*) is listed.
FRU replacement procedures are contained within this chapter of
the manual.
FRU Description
Patient Cable
Keypad Assembly
Top Cover Assembly
LCD Assembly
Printer Assembly
Mainboard Assembly
Power Supply Assembly
Bottom Cover Assembly
Battery Assembly
Real-time Clock (RTC)
MAC™ 800
Basic System FRU Repairs
Visual Inspection
1, 2, 7
3, 6, 7
6, 7
3, 6, 7
6, 7
6, 7
6, 7
6, 7
5
Refer to the Mainboard
Assembly
Functional Checkout
Procedures
1, 2, 3
1, 2, 3, 7
1, 2, 3, 14
1, 2, 3, 6
1, 2, 3, 8
1, 2, 3, 4, 5, 6, 7, 8, 9, 10,
11, 12, 13
1, 2, 3
1, 2, 3
1, 2, 3, 11
Refer to the Mainboard
Assembly
2031504-159D

Hide quick links: